V.Abramov1, N.Akchurin2,
P.Baillon3, A.Ball4, P.de Barbaro5,
V.Barnes6, G.Bayatian7, G.Bencze8,
A.Bodek5, V.Bolotov9, H.Budd5,
L.Dimitrov10, A.Dyshkant1, I.Emelianchik11,
D.Fong4, J.Freeman12, V.Genchev10,
I.Golutvin13, P.Goncharov1, A.Gorin1,
D.Green12, A.Gurzhiev1, E.Guschin9,
V.Hagopian14, P.Iaydjiev10, V.Klimenko9,
Yu.Korneev1, A.Krinitsyn1, V.Kryshkin1,
S.Kunori4, A.Kurilin11, D.Lazic14,
L.Levchuk15, L.Litov16, V.Marin9,
K.Michaud5, Yu.Musienko9, A.Nemashkalo15,
Y.Onel2, P.Petev16, Yu.Petukhov13,
S.Piperov10, V.Podstavkov1, V.Popov15,
A.Proskurjakov9, A.Ronzhin12, A.Rubashkin15,
I.Semenyuk9, Q.Shen6, V.Shmatkov9,
A.Sirunian7, F.Szoncso17, A.Skuja4,
P.Sorokin15, S.Tereschenko1, V.Tyukov13,
T..Virdee3, N.Vlasov13, Volod'ko13,
A.Zaichenko1, A.Zarubin13, A.Zatzerklyany15
1 IHEP, Protvino, RUSSIA
2 University of Iowa, Iowa City, IA, USA
3 CERN, Geneva, SWITZERLAND
4 University of Maryland, College Park, MD, USA
5 University of Rochester, Rochester, NY, USA
6 Purdue University, West Lafayette, IN, USA
7 ErPI, Yerevan, ARMENIA
8 KFKI Research Institute for Particle and Nuclear Physics, Budapest, HUNGARY
9 INR, Moscow, RUSSIA
10 INRNE, Sofia, BULGARIA
11 NCPHEP, Minsk, BELARUS
12 FNAL, Batavia, USA
13 JINR, Dubna, RUSSIA
14 Florida State University, FL, USA
15 KhFTI, Kharkov, UKRAINE
16 SU, Sofia, BULGARIA
17 Institute
fur Hochenergiephysik, Wien, AUSTRIA
The response of tile/fibre calorimeter
to incident muons, electrons, and pions in transverse and longitudinal
magnetic field up to 3 T has been studied. The light yield increase
with magnetic field depends on the field orientation and particle
species. Within the errors the energy resolution and transverse
and longitudinal shower profiles are not affected by magnetic
field.
The CMS hadron calorimeter based on plastic
scintillator read out with wavelength-shifting (WLS) fibre embedded
in the scintillator plate [1] will operate in strong magnetic
field (4 Tesla). Magnetic field changes light yield of scintillator
[2] and affects the shower development [3] that in principle can
depend on the field orientation. For typical collider geometry
the magnetic field is parallel (transverse) to a calorimeter plates
for the central part and perpendicular (longitudinal) to the end
cap. The primary objective of the study was the calorimeter
performance in dependence on magnetic field direction up to
3 T with muon, electron, and pion beams, to identify practical
combinations of thickness and sampling fractions which yield
acceptable resolution for CMS, to test possible candidates of
photodetectors capable to work in magnetic field, to tune up various
hadronic cascade Monte Carlo programs. In section 2 the prototype
calorimeter is described. The experimental set-up is presented
in section 3. The results are discussed in section 4.
A "reconfigurable-stack calorimeter" with brass plates (59 % Cu, 39 % Zn, 1 % Fe, 1 % Mn) as absorber has been built. It can be reconfigured to vary sampling fractions and longitudinal division and, as shown in fig. 1, consists of 3 light tight boxes. Two of them contain stainless steel support frame on which the absorber plates and scintillator plates are hanging. The dimensions of the plates (66 cm x 66 cm) are determined by the size of the hole in the magnet RD5. The third box contains photodetectors.
The active medium is 4 mm thick scintillator, produced in Kharkov, with dimensions 22 cm x 22 cm. A WLS fibre is routed through a key-hole shaped groove milled in the tile. One end of the fibre is machined by flying diamond cutter and aluminised. The 3x3 tiles wrapped in reflective material and fixed between aluminium plates 2 and 1 mm thick form a megatile (fig. 2).
Two configuration of the calorimeter and
the active elements were used.
To a 1 mm diameter WLS fibre (doped with K27, produced in Tver, Russia) a 4 m long PMMA clear fibre is glued in tube to transport light to a photodetector. The WLS fibre turned out to be very fragile and cracked along the bending radius in a month time resulting in drop of light yield. The tile with the fibre is wrapped with aluminised mylar. The fibres from each tower are bundled together, glued in a tube and go into separate holes in a photodetector box which contained 25 mm diameter proximity focused Hybrid Photo Diodes [4]. The HPD outputs are directly connected to preamplifiers and than by 60 m cables to amplifiers to equalise the HPD gain. The HPD HV tension is set to 8.08 kV, the preamplifier gain is 0.125 V /106 e. The gate length is 150 ns corresponding to the pulse length from preamplifiers. Only 12 HPD are available (instead of 36: 9 towers x 4 longitudinal divisions), so the horizontal towers are combined into one.
The calorimeter is divided longitudinally into four compartments with the following sampling: the first two consists of 10 4 cm thick plates, the third 5 8 cm thick absorber plates, and the forth - 2 8 cm thick plates (136 cm of brass).
The light tight box with HPD is placed at
110 cm from the centre of the magnet where the magnetic field
is about 20 % lower than in its maximum. The HPD axes are aligned
along the magnetic field with precision of several degrees. Between
the spills the HPD gain is controlled by single LED, the light
is fanned out by fibres. A PIN diode measures the LED signal to
study the magnetic field dependence on LED light output. The
pedestal position is checked by using random triggers for each
measurements.
To the WLS fibre 0.83 mm diameter (Y11,
produced by Kuraray) a clear fibre is spliced, the other end of
the clear fibre is glued into optical disconnect. The tile with
the fibre is wrapped with Tyvek. A 10 m long optical cable with
disconnects on both ends connects the megatile to a decoder box
containing 18 photo-multiplier (Philips XP 2081 with extended
green photocathode) where fibres are rearranged from layer-to-layer
cables to tower-to-tower bundles. The photo-multipliers have magnetic
shielding. The calorimeter is divided longitudinally into two
read-outs. The sampling is: the first 9 plates 5 cm thick and
the last 11 plates 10 cm thick.
The study of the HCAL performance was carried out on H2 beam at CERN. Beam defining scintillator has dimensions 2 cm x 2 cm. Particle trajectories were measured by drift chambers. The following particles were used: electrons (100 GeV), muons (100 and 300 GeV), and pions (50 - 300 GeV). The magnetic field varied from 0 to 3 T. The deflection of the incident particle by the transverse magnetic field was negligible even for 50 GeV particles. For each measurements about 2x104 triggers were collected.
The calorimeter was installed in the centre of the superconducting RD5 magnet. The magnetic field has nonuniform distribution shown in fig. 3 by thin line.
In the run with longitudinal magnetic field
(see fig. 3b) the magnetic field affected the phototube gain
though the distance between the decoder box and the magnet was
about 7 m. To control stability of PM operation a LED illuminated
all PM through a bundle of fibres. Besides there was a wire radioactive
source [5]. To minimise temperature dependence of the LED for
each particle type magnetic field was changed separately to make
it as quickly as possible.
With rise of magnetic field from 0 to 3 T the gain variation of different HPD was in the range 0-3 %. This dependence was taken into account during data analysis. Position of the LED amplitude distributions was measured for each spill for each channel and the data were corrected.
Neither the calorimeter nor the beam could be moved in horizontal or vertical directions. Therefore only the central tower was illuminated. The calibration constants for towers and longitudinal sections were determined by minimisation of energy resolution.
The magnetic field has a high non-uniformity but it is not clear how to correct the shower profile. The value of the magnetic field cited below corresponds to its meaning at the center of the magnet.
In transverse magnetic field the light yield per scintillator was too small (about 0.2 p.e. per scintillator for minimum ionising particle (m.i.p.)) to reliably measure muon pulse height distribution and as a consequence these dependence was not measured. In the longitudinal magnetic field the WLS fibres were replaced and the light yield was about 1 p.e. per tile for m.i.p. Fig. 4 shows pulse height distribution for muons obtained for the first compartment (9 tiles).
The calorimeter, as was mentioned above, was roughly divided in transverse and longitudinal directions. Comparing normalised distributions we conclude that transverse and longitudinal shower development does not depend on magnetic field for both orientations within the measurement errors and longitudinal division (4 and 2 longitudinal compartments for transverse and longitudinal field correspondingly). The energy resolution also does not depend on magnetic field. Fig. 5 shows the normalized responses vs. transverse magnetic field for 100 GeV electrons and different energy pions. There is almost linear rise of the calorimeter response on magnetic field and dependence on pion energy.
Figs. 6 and 7 show this dependence for
longitudinal magnetic field for electrons and muons and pions.
In this case the behaviour is very close to the scintillator
light yield dependence on magnetic field obtained with radioactive
source and does not depend on particle species and pion energy.
The study of the hadron calorimeter response vs. magnetic field shows that the light yield for muons, electrons, and pions increases up 3 T and for showering particles depends on the field orientation and pion energy. Within the errors there is no appreciable change of transverse and longitudinal shower profile.
There are two effects which lead to calorimeter
light yield dependence on magnetic field: light emission of
scintillators and shower energy absorption in scintillator. Simulations
of the calorimeter response due to only the second effect give
the results presented in the table.
Table. Mote Carlo simulation of calorimeter
respons.
| KT=A(4 T)/A(0 T) | KL = A(4 T)/A(0 T) | |
| 100 GeV electrons | ||
| 100 GeV pions | ||
| 225 GeV pions |
There KT is the ratio of energy absorption in scintillator for 4 T transverse magnetic field over energy absorption in scintillator without magnetic field and KL is the ratio for the longitudinal magnetic field. If the scintillator light yield dependence on magnetic field is added (about 5 % independent on magnetic field orientation) there is a qualitative agreement of the experimental results and the calculation.
For muons if one neglects the electromagnetic interaction at these energies the light yield dependence is mainly determined by the first effect and must not depend on magnetic field orientation.
In conclusion we summarise the main results:
- a calorimeter response vs. magnetic field depends on field orientation and particle species and energy (for transverse field);
-the e/h ratio depends on magnetic field and its orientation;
-a radioactive source calibration to transfer
coefficients obtained with extracted beams and calibration with
muons will be more complicated at high magnetic field so calibration
in situ using physical processes [6] becomes very important.
We gratefully acknowlege the support of
E.Radermacher.
Fig. 1. Schematic view of the "reconfigurable-stack calorimeter".
Fig. 2. The plate with scintillating tile/fibre assembly (megatile).
Fig. 3. Set up of measurements with two field orientation and the magnetic field distribution (thin lines).
Fig. 4. Muon pulse height distribution, the first compartment (9 scintillators).
Fig. 5. The relative change of light yield vs. transverse magnetic field for 100 GeV electrons and 100, 200, and 300 GeV pions.
Fig. 6. The relative change of light yield vs. longitudinal magnetic field for electrons and muons.
Fig. 7. The relative change of light yield vs. longitudinal magnetic field for 50, 100, and 300 GeV pions.
[1] The Compact Muon Solenoid. Letter of Intent. CERN/LHCC 92-3, LHCC/I 1, 1 October 1992.
[2] S.Bertolucci et al., NIM A254 (1987) 561.
D.Blomker et al., IEEE Transactions on Nuclear Science, Vol. 37/2( 1990) 220.
J.P.Cumalat et al., Nucl. Instr. And Meth. A 293 (1990) 606.
[3] V. Abramov, Nucl. Instr. And Meth. A 374 (1996) 34.
[4] H.Arnauden, P.Benetti, L.Boskma et al., CERN-LAA/HC/93-16.
[5] P.de Barbaro et al., Univ. of Rochester preprint UR-1360 (July 1994).
[6] J.Freeman and W.Wu. FNAL-TM-1984 (August
1996).